Technical guide · Updated 2026-11-11
RS-232, RS-485 and Modbus RTU in Industrial Weighing Integration
How to get weight data out of a balance, weighing module or checkweigher and into a PLC, HMI, printer or MES without losing accuracy or traceability.
This guide applies to Definition Electric Ex balance, weighing module and checkweigher installations. Certificate numbers, protection markings, IP ratings and interface details are stated explicitly so they can be verified against the delivered nameplate.
RS-232 on Ex balances
The EX-GB-H, EX-GB-T and EX-GB-G families provide RS-232 ASCII weight output at 1,200 / 2,400 / 4,800 / 9,600 bps, 8 data bits, no parity. It is a simple point-to-point link to a printer, a PC or a data logger - ideal for batch records.
RS-485 on modules and checkweighers
RS-485 supports longer runs and multi-drop networks. The CW checkweigher outputs continuous or on-demand weight data over RS485 using Modbus RTU. The EX-MQ weighing module provides DB9 RS-232 and RS-485 with SBI and Modbus-compatible protocols.
Dual-channel architecture on the strain-gauge module
The strain-gauge high-precision module exposes two independent full-duplex RS485 Modbus RTU channels. Channel 1 has a fixed configuration (station 1, 115200, 8, 1, N) for a local touch-screen HMI; channel 2 is configurable for a host computer or PLC. Both can run simultaneously without bus conflicts.
Reading the register map
Key registers include 4x0000 (Int32 current weight), 4x0002 (decimal places), 4x0003 (unit: 0 = g, 1 = kg, 2 = t), 4x0004 (status bits for stable, zero, zero tracking and sensor OK), 4x0006 (Float32 weight) and 4x0106 (filter time). Multi-register values use little-endian word order (low word first). Supported function codes are 01, 03, 05, 06 and 16.
Integration checklist
Confirm baud rate, parity, stop bits, station ID and byte order before commissioning. Use shielded twisted pair, terminate the bus correctly, keep the weighing cable away from motor drives, and log the calibration weight and zero point with each batch record.
Questions engineers ask
Which interface should I choose for a printer?
RS-232 ASCII output is the simplest link for a ticket or label printer.
Can two masters read the same weigh module?
Use the dual-channel module: one channel for the local HMI, the other for the PLC or host computer.