EX-GB-G Large-Display Explosion-Proof Floor Balance
- 0.001 g Finest readability
- 320 kg Maximum capacity
- IP65 Ingress protection

Typical specifications. Verify model, capacity and approval marking against the instrument nameplate before installation.
Highlights

Enclosure designed to IEC 60079-0 and IEC 60079-1 for Zone 1 and Zone 2 applications.…

Dust-tight enclosure protected against continuous immersion for washdown environments.…

Bright LED display with zero-centre, tare, negative, unit, low-battery and charging status indicators.…

Up to four internal relays (M1 - M4) drive a three-colour HI / OK / LO signal lamp or external equipment.…
Technical data
| Standards | IEC 60079-0 (general requirements); IEC 60079-1 (flameproof enclosure "d") |
|---|---|
| Protection concept | Flameproof "d" - suitable for Zone 1 and Zone 2 |
| Mains supply | AC 220 V (+10% / -15%) / 50 Hz |
| DC adapter | 5 V / 600 mA |
| Battery | 4 V / 4 Ah rechargeable, AC/DC dual use; charge only in a safe area |
| Operating temperature | 0 C to +40 C; storage -25 C to +55 C |
| Relative humidity | <= 85 % RH |
| Warm-up time | 10 - 15 minutes |
| Display | 6-digit bright LED with status annunciators |
| Weighing units | g / kg / lb / oz with unit memory |
| Zero range | Selectable at power-up: 2.7% / 3% / 20% / 50% / 100% FS |
| Filter | Adjustable filter parameter for different environments |
| Overload indication | Display "------" beyond capacity + 9 d |
| Relay outputs | Up to 4 internal relays (M1 - M4), three-colour signal lamp, external 12 V supply |
| Serial output | Optional RS-232C, 1200 / 2400 / 4800 / 9600 bps |
| Warranty | 24 months |
| Model | Max. capacity | Readability (d) |
|---|---|---|
| EX-TCS-30A | 30 kg | 0.5 g |
| EX-TCS-30K | 30 kg | 1 g |
Resources
Customer feedback
Share how this instrument performs in your plant. Every review is emailed to Alvin@definitionscale.com for approval and published only after our team has checked it; your email is never shown.
No reviews have been published yet. Be the first to share your experience with this explosion-proof weighing instrument.
Send us the capacity, readability and area classification. We will confirm the exact model and provide a quotation with lead time.