Calibration and Maintenance of Explosion-Proof Balances

Warm-up, levelling, external calibration, cleaning and safe-area battery charging - a practical routine for keeping an Ex balance accurate.

Key facts

This guide applies to Definition Electric Ex balance, weighing module and checkweigher installations. Certificate numbers, protection markings, IP ratings and interface details are stated explicitly so they can be verified against the delivered nameplate.

Install on a stable, level surface

Precision balances are sensitive to vibration and tilt. Place the instrument on a level, vibration-free bench away from direct sunlight and air-conditioning outlets, and adjust the levelling feet until the indicator is centred.

Allow the specified warm-up

The EX-GB-H needs 60-120 minutes (104-314 class) or 60-90 minutes to reach thermal equilibrium. The EX-GT / EX-TCS indicator needs 10-15 minutes. Skipping warm-up is the most common cause of drift in milligram weighing.

External calibration

Use the supplied calibration weight and the CAL / ZERO key sequence. The EX-MQ module allows a user-definable external weight of roughly one third of maximum capacity. Calibrate after installation, after moving the instrument and after a significant temperature change.

Routine checks

Verify zero at power-up, check repeatability with a known test weight, inspect the glass draft shield and pan, and confirm that RS-232 output still matches the displayed weight. Record the results with the batch documentation.

Cleaning and batteries

Wipe SUS304 surfaces with a damp cloth; do not use high-pressure water on an IP65 enclosure. Never open a flameproof cover while the circuit is energised, and never charge a flameproof battery pack inside the hazardous area - the charger is not explosion-proof designed.

Questions engineers ask

How often should an Ex balance be calibrated?

At installation, after relocation, after a significant temperature change, and at the interval required by your quality system.

Can I open the enclosure to service it?

No. Never open the flameproof cover while the circuit is energised; service must follow the operation manual and the applicable site rules.

Specifications are typical and subject to change. Always verify the approval marking and certificate against your hazardous-area classification before installation.